EVAC Patient Transfer Sheet
A heavy-duty Oxford-weave evacuation and transfer sheet, laboratory-tested for safe horizontal movement of non-ambulatory patients in hospitals, ambulances and care facilities.
Product Overview
Fast, Safe Horizontal Patient Evacuation
The EVAC Patient Transfer Sheet is a finished aid for patient evacuation that helps trained personnel secure and move bedridden or mobility-impaired patients during an emergency evacuation. It is designed to be placed between the mattress and the bed and kept ready for use at all times.
During evacuation, integrated straps secure the patient in place, while reinforced handles and straps allow trained personnel to move them in a controlled way toward the designated safe area. Woven from a heavy-duty Oxford fabric (internally referenced as the Kodro weave) reinforced with PES handles and straps, the sheet is built for repeated hard use.
It can be sprayed and wiped down with hospital-grade disinfectant, or machine washed at temperatures up to 90°C, and is compatible with steam sterilization (autoclave) under validated conditions. Suited to hospitals, nursing homes, clinics and emergency-response environments.

Technical Data
Made in Any Size, On Request
Not suitable for vertical lifting (crane/hoist) — designed for horizontal transfer and repositioning only.
Key Properties
Engineered for Emergency Evacuation
Flame-Retardant
Meets BS 7175:1989 Crib 5 ignition-source acceptance criteria and is rated B1 to DIN 4102-1.
Reinforced Multi-Handle System
Long end handles plus side handles allow secure multi-person lifting with a healthy caregiver posture.
Antibacterial
Meets antibacterial acceptance criteria in independent lab testing.
Autoclave Compatible
Suitable for steam sterilization (autoclave) under validated conditions.
Foldable & Compact
Folds flat for storage between the mattress and bed base — ready for immediate deployment.
250 kg Safe Working Load
Tested with a static load of 500 kg held for 10 minutes with no damage — a 2× safety margin over the rated SWL.
Cleanable & Disinfectable
Machine washable up to 90°C, or spray-and-wipe with hospital-grade disinfectant between patients.
Cold-Flex Resistant
Passed low-temperature bend testing at −20°C with no cracking, per DIN EN 1876-1 — reliable in ambulance and outdoor conditions.
Product Gallery
EVAC Sheet in Use



Evacuation Method
Turn, Slide, Pull — Feet First
Rapid single- or multi-person evacuation
The Evac Sheet stays folded between the mattress and the bed base at all times, ready for immediate use. In an urgent evacuation, one rescuer can move the patient to safety using the foot-end handle alone.
Maintenance
Cleaning, Disinfection & Drying
Inspect after each cleaning for damage, tears, loose seams or damaged handles — do not use if any damage is found. Store clean and dry, protected from direct sunlight, between 0°C and +45°C; do not store under heavy objects.
Independent Load Test — Report 121-2026
Verified to Double the Rated Load
An independent load-capacity test held the sheet under a static test load for 10 minutes with no fabric damage, no seam damage, no handle or strap damage, no permanent deformation, and a passed function check — confirming the sheet’s rated Safe Working Load with a 2× safety margin.
Material Test Report — 100-2026-1
Laboratory-Verified Fabric Performance
| Property | Standard | Requirement | Measured (avg.) | Result |
|---|---|---|---|---|
| Weight | — | 200 g/m² ±10% | 205.4 g/m² | ✓ Pass |
| Tensile Strength (Warp) | TS EN ISO 1421 | ≥ 670 N/5cm | 796.0 N | ✓ Pass |
| Tensile Strength (Weft) | TS EN ISO 1421 | ≥ 700 N/5cm | 874.4 N | ✓ Pass |
| Elongation at Break (Warp) | TS EN ISO 1421 | ≥ 30% | 43.5% | ✓ Pass |
| Elongation at Break (Weft) | TS EN ISO 1421 | ≥ 45% | 62.8% | ✓ Pass |
| Tear Strength (Warp) | EN ISO 4674-1 | ≥ 125 N | 153.0 N | ✓ Pass |
| Tear Strength (Weft) | EN ISO 4674-1 | ≥ 85 N | 96.0 N | ✓ Pass |
| Abrasion Resistance | EN ISO 12947-2 / DIN EN 530 | — | No wear or holes at 50,000 cycles | ✓ Pass |
| Colour Fastness to Rubbing (Wet/Dry) | EN ISO 105-X12 | ≥ 4/5 | 5/5 | ✓ Pass |
| Light Fastness | EN ISO 105-B02 | ≥ 4 | 5 | ✓ Pass |
| Flammability | BS 7175 Crib 5 / DIN 4102-1 | Meets criteria / B1 | Meets criteria | ✓ Pass |
| Antibacterial Activity | — | Meets acceptance criteria | Meets criteria | ✓ Pass |
| Cold-Flex Resistance (−20°C) | DIN EN 1876-1 | No cracking | Passed | ✓ Pass |
Independent lab test report 100-2026-1, dated 07.03.2026, on the standard Navy Kodro fabric used across all sizes.
Regulatory Compliance
Certifications & Standards
Basic UDI-DI 8682550EVAC8W. Manufactured by Gentuğ Tekstil Ürünleri San. ve Tic. A.Ş., Savaş Cad. Söğüt Sok. No:1, 34173 Merter — İstanbul, Türkiye.
Applications
Where the EVAC Sheet Performs
Clinical Use
Indications & Contraindications
Indicated for patients who are:
Non-ambulatory · Semi-conscious or unconscious · Physically immobilized due to injury, illness or surgery · In need of emergency evacuation from a hospital, ambulance, home or disaster zone · At risk of pressure injuries and requiring frequent repositioning.
Not to be used for:
Vertical lifting via crane, hoist or pulley · Loads exceeding the 250 kg safe working load · Use on rough/abrasive surfaces · Use if visibly damaged, torn or with compromised stitching or handles · Long-term bedding — intended for short-term transfer only.
Frequently Asked Questions
What is the maximum weight the EVAC sheet supports?
The EVAC Patient Transfer Sheet has a rated Safe Working Load of 250 kg. An independent load test confirmed it withstands a static test load of 500 kg for 10 minutes with no fabric, seam, or handle damage — a 2x safety margin over the rated load.
Can the sheet be used for vertical lifting?
No. The sheet is designed and tested only for horizontal transfer or repositioning — it must never be used for suspension by crane, hoist or pulley.
How is the sheet cleaned and disinfected between patients?
It’s machine washable up to 90°C, or can be sprayed and wiped with a hospital-grade disinfectant (alcohol- or quaternary-ammonium-based) between patients. Air dry or tumble dry at max 40°C, and it’s compatible with steam sterilization (autoclave) under validated conditions.
How many people are needed to use it safely?
Typically 2-6 trained personnel, depending on the patient’s weight and condition, using all handles evenly to distribute the load. In an urgent evacuation, a single rescuer can pull the patient to safety using the foot-end handle.
How long can the sheet be stored and used?
If stored properly – cool, dry, away from direct sunlight, between 0C and +45C – the sheet can be used for at least 15 years provided it has not been deployed in an actual rescue situation. Regular function checks are recommended.
Request Technical Specifications
For detailed specifications, IFU documents and quotation, reach out to our technical sales team.
Türkçe
Deutsch
Русский