Fabric data, independent barrier test results, wash-durability testing and care instructions for the Caretex Precision reusable operation gown.
Product Overview
The Precision Operation Gown is a reusable surgical gown that pairs two engineered fabrics. Its barrier zones, the front panel and lower sleeves, use a three-layer laminate of polyester microfibre, a breathable, heat-resistant polyurethane membrane and a polyester knit backing. The result is permanently liquid-proof while still letting body moisture escape as vapour.
The rest of the gown is woven polyester with a conductive carbon-yarn grid that dissipates static charge, keeping long operations free of static discomfort. Knit collar and high-temperature-resistant knit cuffs complete the gown.
Every component is chosen to survive industrial laundering and autoclave sterilisation at 134°C. The gown is produced at Gentuğ’s Gerede factory under an EN ISO 13485:2016 quality management system that covers the design and production of operating room gowns.

Garment Construction
Breathable Membrane
The polyurethane membrane in the barrier zones blocks liquids completely, yet lets water vapour from the body pass through and escape to the outside air.
Technical Data
Barrier Performance
| Property | Test Method | Result | Requirement |
|---|---|---|---|
| Resistance to water penetration TZÚ Brno · 2011 |
EN 20811 | ✓ 2,058.6 cm H₂O | ≥ 20 cm H₂O ¹ |
| Hydrostatic pressure Ekoteks · 2024 |
AATCC 127 (Opt. 2) | ✓ > 2,040 cm H₂O | ≥ 50 cm H₂O ² |
| Water impact penetration Ekoteks · 2024 |
AATCC 42 (Opt. 2) | ✓ 0 g | ≤ 1.0 g ² |
| Synthetic blood penetration Shirley Technologies (UKAS) · 2011 |
ASTM F1670 (Proc. B) | ✓ No penetration, 3/3 specimens | No penetration |
| Wet bacterial penetration TZÚ Brno · 2011 |
EN ISO 22610 | ✓ IB 4.6 | ≥ 2.8 ¹ |
| Dry microbial penetration TZÚ Brno · 2011 |
EN ISO 22612 | ✓ 1.3 log CFU (≈ 20 CFU) | ≤ 300 CFU ¹ |
| Microbial cleanliness (bioburden) TZÚ Brno · 2011 |
EN ISO 11737-1 | ✓ 0 log CFU/dm² | ≤ 300 CFU/100 cm² ¹ |
| Bursting strength, dry / wet TZÚ Brno · 2011 |
EN ISO 13938-1 | ✓ 744.7 / 778.5 kPa | ≥ 40 kPa ¹ |
| Tensile strength, dry (MD / CD) TZÚ Brno · 2011 |
EN 29073-3 | ✓ 811.3 / 1,009.3 N | ≥ 20 N ¹ |
| Tensile strength, wet (MD / CD) TZÚ Brno · 2011 |
EN 29073-3 | ✓ 759.7 / 1,032.8 N | ≥ 20 N ¹ |
Results apply to the barrier laminate as tested by independent accredited laboratories (TZÚ Brno, accredited lab No. 1001; Ekoteks, TÜRKAK AB-0583-T; Shirley Technologies, UKAS 1066). ¹ EN 13795-1 requirement, standard performance gown, critical product area. ² AAMI PB70 Level 3 requirement. Full test reports are available on request.
Wash Durability
To confirm reusability, the barrier laminate was put through 90 complete reprocessing cycles by Artea Srl (Italy), an ISO/IEC 17025 testing laboratory. Samples were checked every 10 cycles for tensile strength, thickness, layer adhesion, appearance and colour.
Each cycle:
The laboratory concluded that the material keeps its fitness for use and resists repeated use for at least 90 consecutive cycles. Seams stayed regular throughout and the layers could not be separated in peel testing (ASTM D903).
Care & Reprocessing
Colours
Swatches are indicative; actual fabric shades may vary slightly.
Certifications & Standards
Applications
Product Gallery






The barrier laminate was tested by an ISO/IEC 17025 laboratory through 90 complete reprocessing cycles (wash at 60°C, dry at 60°C, autoclave) with no delamination of the layers, and the laboratory concluded the material remains fit for use for at least 90 consecutive cycles. Actual service life depends on the laundry process and handling.
Conductive carbon yarn is woven into the fabric of the yoke, shoulders, back and hem in a fine grid. It dissipates static charge as it builds up, which removes the discomfort of static electricity during long operations.
The front panel and the lower sleeves, the areas most exposed to blood and fluids, are made from a breathable PU-membrane tri-laminate. This laminate showed no synthetic-blood penetration under ASTM F1670 and passed AAMI PB70 Level 3 barrier testing.
The gown can be industrially washed at up to 90°C and tumble dried, then sterilised by autoclave at 134°C. Do not bleach, iron or dry clean.
Precision is available in blue and green. For size ranges, logo application and volume orders, please contact our technical sales team.